This shows you the differences between two versions of the page.
mgl:functionreferenceeyelink [2012/11/05 14:57] |
mgl:functionreferenceeyelink [2014/12/11 16:37] |
||
---|---|---|---|
Line 38: | Line 38: | ||
If you use the [[:mgl:taskreferenceeyelink|mgl task code]], you do not need to call any of the following functions explicitly. | If you use the [[:mgl:taskreferenceeyelink|mgl task code]], you do not need to call any of the following functions explicitly. | ||
- | ===== mglEyelinkReadEDF ===== | + | ===== mglEyelinkEDFRead ===== |
**purpose:** Reads an Eyelink file into matlab\\ | **purpose:** Reads an Eyelink file into matlab\\ | ||
Line 148: | Line 148: | ||
* A : Auto validation set to the pacing selected in Set Options menu. (Auto trigger ON). EyeLink accepts current fixation if it is stable. | * A : Auto validation set to the pacing selected in Set Options menu. (Auto trigger ON). EyeLink accepts current fixation if it is stable. | ||
* Backspace : Repeats previous calibration target. | * Backspace : Repeats previous calibration target. | ||
+ | == Validation Result == | ||
+ | * GOOD (green background): Errors are acceptable (worst point error < 1.5 deg, avg error < 1.0 deg) | ||
+ | * POOR (red background): Errors are too high for useful eye tracking (worst point error > 2.0 deg, avg error >1.5 deg) | ||
+ | * FAIR (gray background): Errors are moderate, calibration should be improved (any value between the above criteria) | ||
== After Validation == | == After Validation == | ||
* ENTER : Accept validation values | * ENTER : Accept validation values |